The Bradford assay measures protein with Coomassie G-250 dye, which turns from red-brown to blue when it binds protein. Mix 5 µL of sample with 250 µL of dye, wait 5 minutes to 1 hour, read at 595 nm, and convert the reading with a BSA standard curve. The standard assay covers about 125 to 1,000 µg/mL.
The Bradford assay measures how much protein is in a solution by the color change of a dye. It takes a few minutes, needs only a spectrophotometer or plate reader, and works from about 125 to 1,000 µg/mL of protein in its standard form. This guide gives the protocol in all three standard formats with exact volumes, a BSA standard dilution series, a calculator that turns your readings into concentrations, and fixes for readings that go wrong.
Key Facts
- Read at: 595 nm (anywhere from 580 to 610 nm also works)2
- Standard assay range: 125 to 1,000 µg/mL with a BSA standard, 125 to 1,500 µg/mL with gamma-globulin2
- Micro assay range: 1.25 to 10 µg/mL with BSA, for dilute samples2
- Volumes: 5 µL sample + 250 µL dye in a microplate, 20 µL + 1 mL in a cuvette, 100 µL + 5 mL in a tube2
- Timing: at least 5 minutes and no longer than 1 hour at room temperature before reading2
- Main enemies: detergents and basic buffers, which shift the dye’s color on their own2
01 · Principle
How the Bradford assay works
Marion Bradford published the method in 1976.1 It uses Coomassie Brilliant Blue G-250 dissolved in acid. In the acidic reagent the dye sits mostly in its red-brown form, which absorbs light at about 470 nm. When the dye binds a protein it is held in its blue form, which absorbs at 595 nm. The more protein in the sample, the more dye turns blue, and the higher the reading at 595 nm.2
The dye binds mainly to basic amino acids, especially arginine, and to aromatic ones.2 That has two practical consequences. First, different proteins give different amounts of color at the same concentration, so the result is only as accurate as the match between your sample and your standard. Second, very small peptides barely register: the lower limit is a molecular weight of roughly 3,000 to 5,000 daltons.2
Which standard to use
The ideal standard is a purified sample of the protein you are measuring. When you do not have one, you pick a stand-in. Bovine serum albumin (BSA) gives noticeably more color than most proteins, so it suits samples that are mostly albumin. Bovine gamma-globulin gives a response closer to that of a typical protein mixture, such as a cell lysate.2 Either is fine for comparing samples with each other. Whichever you choose, say so when you report the numbers.
02 · Reagents
Bradford reagent and standards
Most labs buy the dye reagent ready to use. It keeps for about a year at 4°C, and it must warm to room temperature and be inverted a few times before use, because cold reagent gives low readings.2
Making Bradford reagent yourself
A widely used recipe:3
- Dissolve the dyeDissolve 50 mg of Coomassie Brilliant Blue G-250 in 50 mL of methanol.3
- Add the acidAdd 100 mL of 85% phosphoric acid to the dye solution.3
- Dilute into waterPour the acid mixture slowly into 850 mL of water and let the dye dissolve completely. Add the acid to the water, never water to the acid.3
- Filter and storeFilter through Whatman No. 1 paper to remove precipitate just before use, and store in a dark bottle at 4°C.3
Methanol and concentrated phosphoric acid both need gloves, eye protection and a fume hood.
BSA standard dilutions
Starting from a 2 mg/mL BSA stock, this series gives enough of each standard to run the microplate assay in triplicate.2 Dilute the standards in the same buffer as your samples, so the buffer affects standards and samples equally.2 If your samples are in phosphate buffered saline, make the standards in PBS too.
| Tube | Final BSA | Take | Add diluent |
|---|---|---|---|
| 1 | 2,000 µg/mL | 20 µL of 2 mg/mL stock | 0 µL |
| 2 | 1,500 µg/mL | 30 µL of 2 mg/mL stock | 10 µL |
| 3 | 1,000 µg/mL | 20 µL of 2 mg/mL stock | 20 µL |
| 4 | 750 µg/mL | 20 µL of tube 2 | 20 µL |
| 5 | 500 µg/mL | 20 µL of tube 3 | 20 µL |
| 6 | 250 µg/mL | 20 µL of tube 5 | 20 µL |
| 7 | 125 µg/mL | 20 µL of tube 6 | 20 µL |
| 8 | 0 (blank) | none | 20 µL |
03 · Protocol
Bradford assay protocol (microplate)
The microplate format is the most common because it uses the least sample and reads every well at once.
What you need
- Bradford dye reagent at room temperature
- BSA or gamma-globulin standard, diluted as in the table above
- Your samples, diluted if needed to fall between 125 and 1,000 µg/mL
- The sample buffer, such as PBS, for the blank and the standard dilutions
- A clear flat-bottom 96-well plate and a plate reader set to 595 nm
- Load the standards and blankPipette 5 µL of each standard and of the blank (buffer only) into separate wells, in duplicate or triplicate.2
- Load the samplesPipette 5 µL of each sample into its own wells, in the same replicate pattern.2
- Add the dyeAdd 250 µL of dye reagent to every well. A multichannel pipette is fastest; pump the plunger a few times to mix, and change tips between rows.2
- IncubateLeave the plate at room temperature for at least 5 minutes. Read within 1 hour, because the color keeps changing slowly.2
- ReadMeasure absorbance at 595 nm.2
- CalculateSubtract the average blank, fit the standard curve and read off the samples, as in section 04 or with the calculator below.2
Other formats
- Plate
- Microplate: 5 µL sample + 250 µL dye. Range 125 to 1,000 µg/mL BSA.2most labs
- 1 mL
- Cuvette: 20 µL sample + 1 mL dye. Same range. Use disposable polystyrene cuvettes; zero the spectrophotometer on the blank.2
- 5 mL
- Tube: 100 µL sample + 5 mL dye. Same range; uses the most reagent.2
- Micro
- Micro assay: equal volumes of sample and dye (1 mL + 1 mL in a cuvette, or 150 µL + 150 µL in a plate) for dilute samples between 1.25 and 10 µg/mL BSA.2
04 · Analysis
Standard curve and calculating concentration
Average the blank wells and subtract that value from every standard and sample. Plot the blank-corrected absorbance of the standards (y axis) against their concentration in µg/mL (x axis), and fit a straight line through the standards in the linear range. Then convert each sample’s corrected absorbance back to a concentration with the line, and multiply by any dilution you made before the assay.2
The curve bends over at the top: past about 1,000 µg/mL of BSA each extra microgram adds less color.2 That is why the 1,500 and 2,000 µg/mL standards are usually left out of the straight-line fit, and why a sample that reads above your highest fitted standard should be diluted and read again rather than extrapolated.
A worked example
These are illustrative readings, the same ones the calculator below opens with. The blank reads 0.45. After subtracting it, the standards from 125 to 1,000 µg/mL give a line of A = 0.000585 × C + 0.025, with R² = 0.998. A sample diluted 1 in 5 reads 0.85, which is 0.40 after the blank. Solving the line: C = (0.40 − 0.025) ÷ 0.000585 = 641 µg/mL in the well. Multiplying by the dilution of 5 gives 3,206 µg/mL, or about 3.2 mg/mL in the original sample.
Calculator
Bradford standard curve calculator
Filled in with example readings so you can see how it works. Replace them with your own 595 nm readings and the line and results update as you type. Nothing you enter leaves this page.
| Fit | BSA µg/mL | A595 (raw) |
|---|---|---|
| 0 (blank) | ||
| Sample | A595 (raw) | Dilution | Protein |
|---|---|---|---|
| - | |||
| - | |||
| - | |||
| - |
Dilution is how many times you diluted the sample before the assay (1 = undiluted, 5 = one part sample in five). A result marked "outside the curve" should be diluted or concentrated and read again, not extrapolated. Example readings are illustrative, not measured data.
05 · Troubleshooting
When Bradford readings go wrong
| What you see | Likely cause and fix |
|---|---|
| Precipitate in the wells | Usually detergent in the sample. Dilute the sample to bring the detergent down, or dialyze it.2 In the standard assay, Bio-Rad lists SDS up to 0.025% and Triton X-100 up to 0.05% as compatible.2 For samples in strong detergent, a detergent-compatible assay such as a modified Lowry is the better tool.2 |
| Everything reads low, standards too | Cold or old dye. Let the reagent reach room temperature before use, and replace it if it is more than a year old.2 |
| Standards fine, samples low | Something in the sample is interfering, or the protein is small. Detergents and basic buffers shift the dye on their own; dilute the sample and make the standards in the same buffer. Proteins below about 3,000 to 5,000 daltons give little color.2 |
| Sample above the top standard | Outside the linear range. Dilute the sample so it falls between 125 and 1,000 µg/mL and read it again.2 |
| Readings drift between plates | Inconsistent timing. Read every plate between 5 minutes and 1 hour after adding dye, and keep standards and samples on the same plate.2 |
| Blue stain in cuvettes | Harmless. Very little dye binds to quartz cuvettes. To clean them, rinse with methanol, or with glassware detergent then water, or soak in 0.1 N HCl for a few hours.2 |
Sources
- Bradford MM. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry 1976;72:248-254. doi:10.1016/0003-2697(76)90527-3
- Bio-Rad Laboratories. Quick Start Bradford Protein Assay instruction manual (4110065A).
- He F. Bradford Protein Assay. Bio-protocol 2011;1(6):e45.